To validate a blood pressure (BP) monitor according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) in a special population of adults with an upper arm circumference of 22 cm or less 37 subjects were recruited. Sphygmomanometer “Armed” identical to previously validated in the general population group YuWell YE660E was used with a “360°” cuff designed for an upper arm circumference of 18 – 36 cm. Sequential measurements of BP were obtained using the test device and the standard mercury sphygmomanometer. A total of 108 sets of comparison data were obtained and analyzed. According to the validation criterion 1, the mean ± SD of the differences was 8.00 ± 6.30 / -16.11 ± 5.15 mmHg. In this special group “Armed” YE660E with “360°” cuff do not meet the requirements of ISO 81060-2:2018.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.