Abstract
To validate a blood pressure (BP) monitor according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) in a special population of adults with an upper arm circumference of 22 cm or less 37 subjects were recruited. The previously validated in general population OMRON M2 Basic (HEM-7121) was used with a HEM CS24 cuff designed for an upper arm circumference of 17–22 cm. Sequential measurements of BP were obtained using the test device and the standard mercury sphygmomanometer. A total of 108 sets of comparison data were obtained and analyzed. According to the validation criterion 1, the mean ± SD of the differences was -7.81 ± 5.20 / -10.66 ± 5.48 mmHg. In this special group HEM-7121 and HEM CS24 do not meet the requirements of ISO 81060-2:2018.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: A.I. Burnasyan Federal Medical Biophysical Center Clinical Bulletin
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.