Abstract

An analysis of main features of unsteady heat flux measuring in shock tubes experiments with a characteristic process time of ~1 μs - 1 ms using sensors based on anisotropic bismuth thermoelements is made. The heat flux behind the reflected shock wave and at the blunt body stagnation point has been measured. Testing of the heat flux calculation method was carried out. The difference between the experimental data and the theoretical value of the heat flux does not exceed 50%. The possible reasons influencing the magnitude of measurement uncertainty are analyzed. The experiments performed have shown the applicability of a sensor based on anisotropic thermoelements and the method for calculating the heat flux from its electrical signal for typical conditions of experiments on shock tubes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.