The analysis of product-related substances and impurities is a critical step in the biopharmaceutical quality control of multiattribute monoclonal antibodies (mAbs), as posttranslational modifications or other variants can influence the product's biological activity. Many approaches are available for variant analysis; however, they are either variant-specific, mAb-specific, time-consuming, or require expensive equipment. Here, we present a generic capillary electrophoretic method based on a neutral-coated capillary which was coupled to mass spectrometry (MS) via the nanoCEasy interface for mAb variant analysis at the subunit level (enzymatically digested and reduced mAb). The method enabled the separation of several (i) size variants (e.g. glycosylation variants) and (ii) charge variants (e.g. c-terminal lysin clipping) as well as (iii) multiple other proteoforms (e.g. additional glycation) and (iv) incompletely reduced subunits. Separated variants were confirmed by MS/MS fragmentation even for small mass deviations like deamidation or open disulfide bridges. The system, initially developed for one mAb, was tested with nine other IgG1s to show the general applicability of the system. The presented multiattribute method enables fast and detailed characterization of mAb variants with little sample preparation and relatively simple separation equipment enabling the separation of a large set of mAb variants.