Intraoperative mapping of the nervous system is used to identify "eloquent" cortical areas. In this technical report, we describe a novel way of mapping the somatosensory cortex so that injury to those critical pathways can be avoided. An 8-year-old female with drug resistant epilepsy presented for resection of a right posterior parietal focal cortical dysplasia. Left median nerve stimulation was used to record somatosensory evoked potentials (SEPs) directly from the somatosensory cortex with a strip electrode. A handheld monopolar electrode was also used to record both the median and tibial SEP. Total intravenous anesthesia with propofol and remifentanil was used. SEP recordings were obtained from a 4-contact strip electrode placed across the central sulcus. A phase reversal was identified and the most likely post central gyrus was noted. With the strip electrode left in place, a monopolar handheld electrode was used to record the median nerve SEPs from different locations on the postcentral gyrus. The tibial nerve was also stimulated to record where the highest amplitude tibial nerve SEP was present. This map was used delineate functionally "eloquent" areas to avoid during surgery. During resection, the median nerve SEP was recorded from the strip electrode continuously. No significant change in the SEP was noted, and the patient awoke without any sensory deficits. Sensory mapping of the cortex is possible with a handheld monopolar electrode. This technique is easy to perform and can help reduce neurological morbidity.