Purpose: We aimed to evaluate the feasibility of using a novel device, the Smart Eye Camera (SEC), for assessing tear meniscus height (TMH) after fluorescein staining and the agreement of the results with measurements obtained using standard slit lamp examination. Methods: TMH was assessed using both SEC and conventional slit lamp examination. The images were analyzed using the software ImageJ 1.53t (National Institutes of Health, Bethesda, MD, USA). A common measurement unit scale was established based on a paper strip, which was used as a calibration marker to convert pixels into metric scale. A color threshold was applied using uniform parameters for brightness, saturation, and hue. The images were then binarized to black and white to enhance the representation of the tear menisci. A 2 mm area around the upper and lower meniscus in the central eye lid zone was selected and magnified 3200 times to facilitate manual measurement. The values obtained using SEC were compared with those obtained with a slit lamp. Results: The upper and lower TMH values measured using the SEC were not statistically different from those obtained with a slit lamp (0.209 ± 0.073 mm vs. 0.235 ± 0.085, p = 0.073, and 0.297 ± 0.168 vs. 0.260 ± 0.173, p = 0.275, respectively). The results of Bland–Altman analysis demonstrated strong agreement between the two instruments, with a mean bias of −0.016 mm (agreement limits: −0.117 to 0.145 mm) for upper TMH and 0.031 mm (agreement limits: −0.306 to 0.368 mm) for lower TMH. Conclusions: The SEC demonstrated sufficient validity and reliability for assessing TMH in healthy eyes in a clinical setting, demonstrating concordance with the conventional slit lamp examination.