BackgroundCoronary bifurcation lesions (CBL) comprise 15 %–20 % of percutaneous coronary procedures and remain a challenge despite advances in stent and interventional techniques. The BioMime™ Branch sirolimus-eluting coronary side-branch stent (BBSES) is specifically designed for CBL treatment in conjunction with a standard drug-eluting stent (DES). We report the first single-center experience of treating complex CBL with the novel BBSES. MethodsThis is a retrospective, single-center study involving consecutive prospectively identified patients who underwent treatment of true CBL with the BBSES. The protocol included BBSES+DES implantation in the CBL and simultaneous final kissing balloon inflation. ResultsFifty-eight CBL were treated in 58 consecutive patients (89.6 % men, mean age 69.0 ± 9.5 years) presenting primarily with stable angina (84.4 %) and true (Medina 1,1,1,) CBL. Procedural success was 100 % without major adverse cardiac events (MACE). At a median follow-up of 18 months, one sudden death was reported that was accounted as possible late stent thrombosis. One patient had spontaneous myocardial infarction due to subacute thrombosis of a DES implanted in the main vessel proximally to the BBSES before the index procedure. Another patient was hospitalized for atrial fibrillation. ConclusionsThis is the first clinical experience to date of true CBL treatment with the BBSES demonstrating high procedural success, no in-hospital MACE and sustained clinical results at a median follow-up of 18 months.