Contouring accuracy is critical in modern radiotherapy. Several tools are available to assist clinicians in this task. This study aims to evaluate the performance of the smoothing tool in the ARIA system to obtain more consistent volumes. Eleven different geometric shapes were delineated in ARIA v15.6 (Sphere, Cube, Square Prism, Six-Pointed Star Prism, Arrow Prism, And Cylinder and the respective volumes at 45° of axis deviation (_45)) in 1, 3, 5, 7, and 10cm side or diameter each. Post-processing drawing tools to smooth those first-generated volumes were applied in different options (2D-ALL vs 3D) and grades (1, 3, 5, 10, 15, and 20). These volumetric transformations were analyzed by comparing different parameters: volume changes, center of mass, and DICE similarity coefficient index. Then we studied how smoothing affected two different volumes in a head and neck cancer patient: a single rounded node and the volume delineating cervical nodal areas. No changes in data were found between 2D-ALL or 3D smoothing. Minimum deviations were found (range from 0 to 0.45cm) in the center of mass. Volumes and the DICE index decreased as the degree of smoothing increased. Some discrepancies were found, especially in figures with cleft and spikes that behave differently. In the clinical case, smoothing should be applied only once throughout the target delineation process, preferably in the largest volume (PTV) to minimize errors. Smoothing is a good tool to reduce artifacts due to the manual delineation of radiotherapy volumes. The resulting volumes must be always carefully reviewed.