BackgroundThe latest-generation Evolut FX TAVR system (Medtronic) offers several potential design improvements over its predecessors, but early reported experience has been limited. ObjectivesThis study sought to report our multicenter, limited market release, first-in-human experience of transcatheter aortic valve replacement (TAVR) with the Evolut FX system and compare it with a single-center PRO+ (Medtronic) experience. MethodsFrom June 27 to September 16, 2022, 226 consecutive patients from 9 US centers underwent transfemoral TAVR with the Evolut FX system for native aortic stenosis (89.4%) or prosthetic valve degeneration (10.6%). Commissural alignment was defined as 0° to 30° between native and FX commissures. Patient, anatomical, and procedural characteristics were retrospectively reviewed, and 30-day clinical and echocardiographic outcomes per Valve Academic Research Consortium-3 definitions were reported. ResultsOf 226 patients, 34.1% were low risk, 4% had a bicuspid valve, and 11.5% had a horizontal root (≥60°). Direct Inline sheath (Medtronic) was used in 67.6% and Lunderquist stiff wire (Cook Medical) in 35.4% of cases. Optimal hat marker orientation during deployment was achieved in 98.4%, with commissural alignment in 96.5%. At 30 days, 14.3% mild, 0.9% moderate, and no severe paravalvular leak were observed. Compared with the Evolut PRO+ experience from 1 center, FX had a more symmetrical implantation with shallower depth at the left coronary cusp (P < 0.001), fewer device recaptures (26.1% vs 39.5%; P = 0.004), and improved commissural alignment (96.5% vs 80.2%; P < 0.001). ConclusionsThe Evolut FX system demonstrated favorable 30-day outcomes with a significant improvement over PRO+ in achieving commissural alignment, fewer device recaptures, and more symmetrical implantation. These features may benefit younger patients undergoing TAVR with the supra-annular, self-expanding valve, where lifetime management would be important.