This prospective study aimed to investigate the ability of cardiac autonomic nervous system (CANS) activity assessment to predict appropriate implantable cardioverter-defibrillator (ICD) therapy in patients with coronary artery disease (CAD) during long-term follow-up period. We enrolled patients with CAD and ICD implantation indications that included both secondary and primary prevention of sudden cardiac death. Before ICD implantation CANS was assessed by using heart rate variability (HRV), myocardium scintigraphy with 123I-meta-iodobenzylguanidine (123I-MIBG) and erythrocyte membranes β-adrenoreactivity (EMA). The study's primary endpoint was the documentation of appropriate ICD therapy. Of 45 (100.0%) patients, 15 (33.3%) had appropriate ICD therapy during 36 months follow-up period. Patients with appropriate ICD therapy were likely to have a higher summed 123I-MIBG score delayed (p < 0.001) and lower 123I-MIBG washout rate (p = 0.008) indicators. These parameters were independently associated with endpoint in univariable and multivariable logistic regression. We created a logistic equation and calculated a cut-off value. The resulting ROC curve revealed a discriminative ability with AUC of 0.933 (95% confidence interval 0.817-0.986; sensitivity 100.00%; specificity 93.33%). Combined CANS activity assessment is useful in prediction of appropriate ICD therapy in patients with CAD during long-term follow-up period after device implantation.
Read full abstract