Description of Cases: Case 1: A 72 year old male with atrial fibrillation (AF) and tachy-brady syndrome underwent Micra implantation. On post-procedure day 2, pacing thresholds were unacceptably elevated. A new Micra device was deployed; subsequently, a microsnare was placed through the Micra delivery catheter and the docking button of the old Micra was successfully snared (Figure A). The old Micra was then captured over the delivery catheter and removed from the body. Case 2: A 76 year old male underwent Micra implantation and AV node ablation for AF with persistent tachycardia. One month later, he presented with heart failure and interval decrease in ejection fraction from 66% to 33%. After biventricular leads were implanted, an Agilis catheter was advanced towards the Micra. The docking button was successfully snared with a gooseneck snare (Figure B), and the Micra was extracted with gentle traction. Case 3: A 76 year old male with complete heart block and longstanding persistent AF received a Micra. Nine months later, he presented with heart failure and interval decrease in ejection fraction from 55% to 25%. A defibrillator lead was implanted but no suitable CS branch was identified. The Aveir retrieval catheter was advanced and the Micra docking button was captured with the triloop snare (Figure C). The protective sleeve was advanced over the Micra and it was successfully removed. Subsequently, a left bundle area lead was placed. Discussion: Although uncommon, extraction of the tined leadless pacemaker (Micra) may be indicated at times. As rates of implantation continue to increase, establishing uniform procedures for extraction becomes increasingly important. These three cases highlight unique options for Micra extraction, and the feasibility of each approach. However, a dedicated tool and protocol would be ideal for patient safety and consistency. No complications occurred with the above cases.
Read full abstract