Academics are far from a consensus regarding the effects of pneumatic tube system (PTS) delivery on sample integrity and laboratory test results. As for the reasons for conflicting opinions, each PTS is uniquely designed, sample tubes and patient characteristics differ among studies. This study aims to validate the PTS utilized in Ankara City Hospital for routine chemistry, coagulation, and hematology tests by comparing samples delivered via PTS and porter. The study comprises 50 healthy volunteers. Blood samples were drawn into three biochemistry, two coagulation, and two hemogram tubes from each participant. Each of the duplicate samples was transferred to the emergency laboratory via Swiss log PTS (aka PTS-immediately) or by a porter. The last of the biochemistry tubes were delivered via the PTS, upon completion of coagulation of the blood (aka PTS-after). The results of the analysis in these groups were compared with multiple statistical analyses. The study did not reveal any correlation between the PTS and serum hemolysis index. There were statistically significant differences in several biochemistry tests. However, none of them reached the clinical significance threshold. Basophil and large unidentified cell (LUC) tests had poor correlations (r=0.47 and r=0.60; respectively) and reached clinical significance threshold (the average percentages of bias, 10.2%, and 15.4%, respectively). The remainder of the hematology and coagulation parameters did not reach clinical significance level either. The modern PTS validated in this study is safe for sample transportation for routine chemistry, coagulation, and hematology tests frequently requested in healthy individuals except for basophil and LUC.