Brachytherapy (BT), especially in high dose rate (HDR), has become increasingly complex owing to the use of image-guided techniques and the introduction of advanced applicators. Consequently, radiotherapy technologists and medical physicists (RTMPs) require substantial training to enhance their knowledge and technical skills in image-guided brachytherapy. However, the current status of the RTMP workload, individual abilities and quality control (QC) of BT units in Japan remains unclear. To address this issue, we conducted a questionnaire survey from June to August 2022 in all 837 radiation treatment facilities in Japan involving RTMPs. This survey focused on gynecological cancers treated with HDR-BT (GY-HDR) and permanent prostate implantation using low-dose-rate BT (PR-LDR). The responses revealed that the average working time in the overall process for HDR varied: 120min for intracavitary BT and 180min for intracavitary BT combined with interstitial BT. The QC implementation rate, in accordance with domestic guidelines, was 65% for GY-HDR and 44% for PR-LDR, which was lower than the 69% observed for external beam radiation therapy (EBRT). Additionally, the implementation rate during regular working hours was low. Even among RTMP working in facilities performing BT, the proportion of those able to perform QC for BT units was ~30% for GY-HDR and <20% for PR-LDR, significantly lower than the 80% achieved for EBRT. This study highlights the vulnerabilities of Japan's BT unit QC implementation structure. Addressing these issues requires appropriate training of the RTMP staff to safely perform BT tasks and improvements in practical education and training systems.