Electrogram pattern recognition by way of morphologic analysis has been proposed as a technique that may improve discrimination of ventricular tachycardia from sinus rhythm by anti-tachycardia devices. The potential impact that increases in heart rate and sympathetic tone could have on such techniques, however, has not been previously determined. A comparative study was undertaken to quantify possible changes in ventricular electrogram morphology using correlation waveform, area of difference, and amplitude analyses in 6 patients during atrial overdrive pacing at cycle lengths of 600 and 400 ms (group A), in 13 patients during infusions of physiologic doses of epinephrine (group B), and in 20 patients undergoing infusions of isoproterenol (group C). Four patients were in both groups A and B. A bipolar intraventricular template of cardiac depolarization during sinus rhythm at rest was compared with depolarization during subsequent passages of sinus rhythm at rest and subsequently increased heart rate. In 36 of 39 patients, waveform configuration as assessed by correlation waveform analysis remained relatively stable during atrial overdrive pacing, epinephrine infusion, and isoproterenol infusion when compared with sinus rhythm at rest. The correlation value did not fall below 0.950 in any patient. Area of difference values for the same 36 patients changed by an average of 6 and 37% during atrial overdrive pacing at cycle lengths of 600 and 400 ms intervals, respectively, by 3% during epinephrine infusion, and by 17% during isoproterenol infusion. In these same patients, there was an average change in electrogram amplitude of −1% during atrial overdrive pacing at 600 ms, 26% during pacing at 400 ms, −1% during epinephrine infusion, and 12% during isoproterenol infusion. Little or no correlation was found between relative changes of amplitude and changes of correlation coefficient during any intervention. However, changes in amplitude and changes in area of difference exhibited weak to moderate correlation during infusion of epinephrine and isoproterenol. Despite some amplitude variation, overall waveform morphology remains relatively constant during increases in heart rate with or without concurrent changes in sympathetic tone. These findings suggest that heart rate increases are unlikely to confound time-domain detection schemes used in antitachycardia devices.
Read full abstract