Purpose:To evaluate corneal ap thickness with Moria M2 microkeratome in Laser in situ keratomileusis (LASIK) using Anterior segment optical coherence tomography (AS-OCT). Methods: 170 eyes of 85 patients underwent LASIK using VISX Star S4 Excimer Laser System. The Moria M2 single-use head 90m microkeratome was used to create nasal hinged corneal aps. The right eye (RE) was operated rst followed by the left eye (LE), using the same blade. Flap thickness was analyzed using AS-OCT. Statistical Analysis: SPSS Software for Windows (version 18.0, SPSS, Inc). Results: The mean spherical equivalent (MSE) was - 4.17 diopters (D) + 2.5(range: -12.5 to + 2.375 D), preoperative keratometry was 44.02 +1.58D (range 40.19-48.95D), and corneal thickness was 535.62 ± 34.6 m (468-661m) in all eyes. The mean corneal ap thickness was 117.80 ± 9.12m (96 -138m), 112.98 ± 9.86m (93-142m), and 115.35 ± 9.7m (93-142m) in the RE, LE, and both eyes respectively. There was a negative signicant relationship between MSE and the ap thickness (p<0.001). Females had thinner aps as compared to males (p=0.013). There was a weak positive correlation between the pre-operative pachymetry and ap thickness (p=0.039). No signicant relationship was found between preoperative keratometry, age and ap thickness. RE had a thicker ap than the LE which was statistically signicant (p<0.001). Conclusions:The aps created using Moria M2 90 microkeratome were thicker than intended and reproducible. Preoperative keratometry and age did not inuence ap formation. Thick corneas had thicker aps and vice versa. Higher refractive errors and re-use of blades cause thinner aps.