Patulin is one of the mycotoxins frequently detected in apples and derivatives, representing a major food safety risk. This study aimed to validate a high-performance liquid chromatography (HPLC) method with an ultraviolet (UV) detector for patulin quantification and assess its occurrence in apple beverages marketed in Morocco. The validation parameters showed satisfactory results with adequate linearity (R > 0.997), a relative standard deviation below 2.5%, repeatability between 3.6 and 7.1%, reproducibility between 3.9 and 11.5%, a limit of quantification (LOQ) of 4 μg/L, and recoveries close to 100% for three levels. Analysis of 30 samples revealed patulin levels ranging from 0 to 16.36 μg/L, with 50% of samples showing negative levels. All positive results remained below the regulatory maximum limit of 50 μg/L. These findings affirm the efficacy of the HPLC proposed method in ensuring compliance with patulin regulations in apple beverages, underlining its importance in safeguarding food safety.