Detection of slight pH changes in skin interstitial fluid (ISF) is crucial yet challenging for studying pathological processes and understanding personal health conditions. In this work, we construct an i-motif DNA based fluorescent ratiometric microneedle sensing patch (IFR-pH MN patch) strategy that enables minimally invasive, high-resolution, and sensitive transdermal monitoring of small pH variations in ISF. The IFR-pH MN patch with advanced integration of both ISF sampling and pH sensing was fabricated from the cross-linking of gelatin methacryloyl and methacrylated hyaluronic acid, wrapping with pH-sensitive hairpin-containing i-motif DNA based fluorescent ratiometric probes in the matrix. Because it is mechanically robust for skin penetration and has high swelling ability, the IFR-pH MN patch could be quickly extracted as sufficient liquid from agarose gel (∼56.4 μL in 10 min). Benefiting from conformation changes of the hairpin-containing i-motif DNA under pH variation and ratiometric fluorescence signal readout, the IFR-pH MN patch could quantitate pH over a small range between pH 6.2 and 6.9 with an accuracy of 0.2 pH units in the mimic skin model. Furthermore, in vivo testing on wound and tumor mouse models indicated the ability of the biocompatible IFR-pH MN patch to penetrate the skin for obtaining transdermal pH values, demonstrating the potential applications in monitoring and intervention of pathological states.
Read full abstract