Harmonic motion imaging (HMI) is an ultrasound elastography technique that estimates the viscoelastic properties of tissues by inducing localized oscillatory motion using focused ultrasound (FUS). The resulting displacement, assumed to be inversely proportional to the tissue local stiffness, is estimated using an imaging array based on RF speckle tracking. In conventional HMI, this is accomplished with plane-wave (PW) imaging, which inherently suffers from low lateral resolution. Coherent PW compounding (PWC) leverages spatial and temporal resolution using synthetic focusing in transmit. In this study, we introduced focused imaging with parallel tracking in HMI and compared parallel tracking of various transmit F-numbers (F/2.6, 3, 4, and 5) qualitatively and quantitatively with PW and PWC imaging at various compounded angle ranges (6°, 12°, and 18°). An in silico model of a 56-kPa spherical inclusion (diameter: 3.6 mm) embedded in a 5.3-kPa background and a 5.3-kPa elastic phantom with cylindrical inclusions (Young's moduli: 22-56 kPa, diameters: 2.0-8.6 mm) were imaged to assess different tracking beam sequences. Speckle biasing in displacement estimation associated with parallel tracking was also investigated and concluded to be negligible in HMI. Parallel tracking in receive (Rx) resulted in 2%-7% and 8%-12% increase compared to PW imaging ( ) in HMI contrast and contrast-to-noise ratio in silico and phantoms. Focused imaging with parallel tracking in Rx was concluded to be most robust among PW and PWC imaging for displacement estimation, and its preclinical feasibility was demonstrated in postsurgical human cancerous breast tissue specimens and in vivo murine models of breast cancer.