Purpose: To evaluate and compare the deliverability of ‘sawtooth’ proton arc therapy (PAT) plans relative to static intensity modulated proton therapy (IMPT) at a cyclotron-based clinical facility. Methods: The delivery of single and dual arc Sawtooth PAT plans for an abdominal CT phantom and multiple clinical cases of brain, head and neck (H&N) and base of skull (BoS) targets was emulated under the step-and-shoot and continuous PAT delivery regimes and compared to that of a corresponding static IMPT plan. Results: Continuous PAT delivery increased the time associated with beam delivery and gantry movement in single/dual PAT plans by 4.86/7.34 min (brain), 7.51/12.40 min (BoS) and 6.59/10.57 min (H&N) on average relative to static IMPT. Step-and-shoot PAT increased this delivery time further by 4.79 min on average as the delivery was limited by gantry motion. Conclusions: The emulator can approximately model clinical sawtooth PAT delivery but requires experimental validation. No clear benefit was observed regarding beam-on time for sawtooth PAT relative to static IMPT.