Until now, giant intracranial aneurysms (GIAs) have in many cases been a vascular disease that was difficult or impossible to treat, not least due to the lack of availability of a large-format stent. In this multicentre study, we report on the first five clinical applications of the Accero®-Rex-Stents (Acandis, Pforzheim, Germany) in the successful treatment of fusiform cerebral giant aneurysms. The Accero®-Rex-Stents are self-expanding, braided, fully radiopaque Nitinol stents designed for aneurysm treatment. The stent is available in three different sizes (diameter 7-10 mm, length 30-60 mm) and intended for endovascular implantation in vessels with diameters of 5.5-10 mm. Five patients (all male, age 54.4 ± 8.1 years) with large fusiform aneurysms of the posterior circulation were treated endovascularly using the Accero®-Rex-Stents. There were no technical complications. One major ischemic complication occurred. A significant remodeling and reduction in the size of the stent-covered aneurysms was already seen in the short-term post-interventional course. The Accero®-Rex-Stents were successfully and safely implanted in all five patients with fusiform giant aneurysms, showing technical feasibility with promising initial results and significant aneurysm size reduction in already available follow-up imaging. With the Accero-Rex-Stents, a new device is available that offers another treatment option for rare cerebral fusiform giant aneurysms with very large parent vessels.