The objective of this pilot study was to assess the reliability of superb microvascular ultrasound (SMI) for the measurement of the cerebrospinal fluid (CSF) flow within VPS systems as an indirect sign for shunt dysfunction. Asymptomatic hydrocephalus patients, with a VPS system implanted between 2017 and 2021, were prospectively enrolled in the study. Using SMI, the CSF flow within the proximal and distal catheters were analysed. Before and after pumping the shunt reservoir, intraabdominal free fluid, optical nerve sheath diameter (ONSD), and papilla diameter (PD) were evaluated and correlated with the amount of valve activation. Nineteen patients were included. A flow was detectable in 100% (N = 19) patients in the proximal and in 89.5% (N = 17) in the distal catheter. The distal catheter tip was detectable in 27.7% (N = 5) patients. Free intraabdominal fluid was initially detected in 21.4% (N = 4) patients and in 57.9% (N = 11) at the end of the examination (P = 0.049). ONSD was significantly lower after pump activation (4.4 ± 0.9 mm versus 4.1 ± 0.8 mm, P = 0.049). Both peak velocity and flow volume per second were higher in proximal compared to distal catheters (32.2 ± 45.2 versus 5.6 ± 3.7 cm/sec, P = 0.015; 16.6 ± 9.5 ml/sec versus 5.1 ± 4.0 ml/sec, P = 0.001, respectively). No correlation was found between the number of pump activations and the changes in ONSD (P = 0.975) or PD (P = 0.820). SMI appears to be a very promising non-invasive diagnostic tool to assess CSF flow within the VPS systems and therefore affirm their function. Furthermore, appearance of free intraperitoneal fluid followed by repeated compression of a shunt reservoir indicates an intact functioning shunt system.