Backgrounds With respect to the knife's design in colorectal endoscopic submucosal dissection (ESD), diameter, water jet function, and electric power are important because these relate to efficient dissection. In this study, we analyzed a novel, narrow ball tip-typed ESD knife with water jet function (Flush knife BT-S, diameter: 2.2 mm, length: 2000 mm, Fujifilm Co., Tokyo, Japan) compared to a regular diameter knife (Flush knife BT, diameter: 2.6 mm, length: 1800 mm). Methods In laboratory and clinical research, electric power, knife insertion time, vacuum/suction amount with knife in the endoscopic channel, and water jet function were analyzed. We used a knife 2.0 mm long for BT-S and BT knives. Results The BT-S showed faster mean knife insertion time (sec) and better vacuum amount (ml/min) compared to the BT (insertion time: 16.7 versus 21.6, p < 0.001, vacuum amount: 38.0 versus 14.0, p < 0.01). Additionally, the water jet function of the BT-S was not inferior. In 39 colorectal ESD cases in two institutions, there were mean 4.7 times (range: 1–28) of knife insertion. Suction under knife happened 59% (23/39) and suction of fluid could be done in 100%. Conclusions Our study showed that the narrow knife allows significantly faster knife insertion, better vacuum function, and effective clinical results.