The ion motion in a quadrupole ion trap of hyperbolic geometry is well described by the Mathieu equations. A simpler cylindrical ion trap has also gained significance and has been used by us for fluorescence-spectroscopy experiments. This design allows for the easy replacement of the end-cap with a mesh, enhancing the photon collection. It is crucial to obtain a firm understanding of the ion motion in cylindrical ion traps and their capability as mass spectrometers. We present here an empirical method of calibrating a cylindrical ion trap based on fluorescence detection. This can be done nearly background-free in a pulsed experiment. The ions are located at the center of the trap, where the field is primarily quadrupolar, and here an effective Mathieu description is found through an effective geometry parameter. In spectroscopy experiments, high buffer-gas pressures are needed to efficiently cool the ions, which complicates the ions' motion and hence their stability. Still, simulations show that the stability diagram closely aligns with the Mathieu diagram, albeit shifted due to collisions. We map the stability diagram for six molecular ions by fluorescence collection from four cations and two anions spanning m/z from 212 to 647. The stability diagram is parametrized through the Mathieu functions with an m/z-dependent effective geometry parameter and a q-dependent shrinkage of the diagram. Based on the calibration, we estimate the mass resolution to be +7/-3 Da for ions with masses in the hundreds of Da.