Stent misdeployment (SMD) is a feared technical challenge of EUS-guided choledochoduodenostomy using a lumen apposing stent (EUS-CDS) that has been poorly characterized. We aim to ascertain the rate of SMD in EUS-CDS for malignant distal biliary obstruction (MDBO) and describe its outcomes while identifying variables associated with its occurrence. In addition, we aim to propose a de novo SMD classification. A post hoc analysis of two RCTs comparing EUS-CDS vs. endoscopic retrograde cholangiopancreatography in MDBO. The main endpoint was the rate of SMD classified as misdeployment of the distal flange (type I), proximal flange (type II), contralateral bile duct wall injury (type III), or double mucosal puncture (type IV). A multivariable analysis was performed to identify variables associated with the odds of SMD and/or technical failure and clinical failure or stent dysfunction. 152 patients were included with a technical success of 93.4%. SMDs occurred in 11 (7.2%) patients (95% CI, 3.1%-11.4%): 8 type I, 1 type II, and 2 type III. Endoscopic salvage was successful in 81.8%. SMDs led to an adverse event (AE) in 4 cases (2 mild and 2 moderate) with an overall AE rate of 2.6% (95% CI, 0.7%-6.6%). On multivariable analysis, an extrahepatic bile duct diameter of ≤ 15 mm was associated with increase odds of SMD and/or technical failure. SMD is relatively common in EUS-CDS and is associated with an extra-hepatic bile duct diameter measuring ≤ 15 mm. The majority of SMDs can be rescued endoscopically with low risk for AE.
Read full abstract