Atrial fibrillation (AFib) is the most common sustained heart arrhythmia and is characterized by irregular and excessively frequent ventricular contractions. Early diagnosis of AFib is a key step in the prevention of stroke and heart failure. In this paper, we present a comprehensive time–frequency pattern analysis approach for automated detection of AFib from smartphone-derived seismocardiography (SCG) and gyrocardiography (GCG) signals. We sought to assess the diagnostic performance of a smartphone mechanocardiogram (MCG) by considering joint SCG–GCG recordings from 435 subjects including 190 AFib and 245 sinus rhythm cases. A fully automated AFib detection algorithm consisting of various signal processing and multidisciplinary feature engineering techniques was developed and evaluated through a large set of cross-validation (CV) data including 300 (AFib = 150) cardiac patients. The trained model was further tested on an unseen set of recordings including 135 (AFib = 40) subjects considered as cross-database (CD). The experimental results showed accuracy, sensitivity, and specificity of approximately 97%, 99%, and 95% for the CV study and up to 95%, 93%, and 97% for the CD test, respectively. The F1 scores were 97% and 96% for the CV and CD, respectively. A positive predictive value of approximately 95% and 92% was obtained, respectively, for the validation and test sets suggesting high reproducibility and repeatability for mobile AFib detection. Moreover, the kappa coefficient of the method was 0.94 indicating a near-perfect agreement in rhythm classification between the smartphone algorithm and visual interpretation of telemetry recordings. The results support the feasibility of self-monitoring via easy-to-use and accessible MCGs.