Filamentous fungi are an emergent cause of severe infections in immunocompromised patients. Timely and accurate identification is crucial to initiate appropriate therapy. Traditional identification methods are time-consuming, labor-intensive, and operator-dependent. Matrix-assisted laser desorption/ionization-time-of-flight (MALDI-TOF) mass spectrometry is a rapid and easy-to-perform identification method. The effectiveness of a commercial MALDI-TOF MS platform to identify filamentous fungi in a clinical laboratory was evaluated. The study included 67 fungal isolates from 35 species/species complexes, which were identified and confirmed in mycology reference laboratories; 32 derived from clinical samples, 34 from strain collections and one was an ATCC strain. The study used the VITEK® MS system (v3.2.0 database), after sample extraction by VITEK® MS Mould Kit. Results were classified as “correct species”, “correct species complex”, “correct genus”and “no identification”. VITEK® MS correctly identified 91.0% of isolates (58.2% to species, 29.9% to species complex, and 1.5% to genus level only). In 82%, the result matched the species/species complex identified by reference methods. No misidentifications were observed. The kit was rapid and easy to use. In conclusion, the VITEK® MS system showed a high capability to accurately identify filamentous fungi in a clinical laboratory.