Left ventricular assist device (LVAD) malposition has been linked to hemocompatibility-related adverse events (HRAEs). This study aimed to identify preoperative anatomical landmarks and postoperative pump position, associated with HRAEs during LVAD support. Pre- and postoperative chest X-ray measures (≤14 days pre-implantation, first postoperative standing, 6, 12, 18, and 24 months post-implantation) were analyzed for their association with HRAEs over 24 months in 33 HeartMate 3 (HM3) patients (15.2% female, age 66 (9.5) years). HM3 patients with any HRAE showed significantly lower preoperative distances between left ventricle and thoracic outline (dLVT) (25.3 ± 10.2 mm vs. 40.3 ± 15.5 mm, p = 0.004). A ROC-derived cutoff dLVT ≤ 29.2 mm provided 85.7% sensitivity and 72.2% specificity predicting any HRAE during HM3 support (76.2% (>29.2 mm) vs. 16.7% (≤29.2 mm) freedom from HRAE, p < 0.001) and significant differences in cardiothoracic ratio (0.58 ± 0.04 vs. 0.62 ± 0.04, p = 0.045). Postoperative X-rays indicated lower pump depths in patients with ischemic strokes (9.1 ± 16.2 mm vs. 38.0 ± 18.5 mm, p = 0.007), reduced freedom from any neurological event (pump depth ≤ 28.7 mm: 45.5% vs. 94.1%, p = 0.004), and a significant correlation between pump depth and inflow cannula angle (r = 0.66, p < 0.001). Longitudinal changes were observed in heart-pump width (F(4,60) = 5.61, p < 0.001). Preoperative X-ray markers are associated with postoperative HRAE occurrence. Applying this knowledge in clinical practice may enhance risk stratification, guide therapy optimization, and improve HM3 recipient management.