ObjectiveEvaluation of the mid-long-term kinematics of single-level Bryan artificial cervical disc replacement (ACDR) in vivo by analyzing the center of rotation (COR) at the operated level.MethodsA retrospective analysis was conducted using data collected from 38 patients who underwent single-level Bryan ACDR from January 2010 to March 2013. Radiological parameters including range of motion (ROM), lordosis angle, translation, and COR were obtained. Clinical outcomes were assessed based on Odom Criteria, modified Japanese Orthopedic Association (mJOA), Neck Disability Index (NDI), and Visual Analogue Scale (VAS) scores. Correlations between COR and other follow-up data were discussed at the last follow-up.ResultsCompared with preoperative values, the last follow-up data showed that 86.84% of cases achieved good-or-excellent outcomes based on Odom criteria; Significant improvements were observed across all scales assessed for clinical outcomes (P < 0.05); Lordosis angle was significantly increased in both the overall cervical spine and the operated level (P < 0.05); ROM of the overall cervical spine, operated level, and adjacent levels was preserved (P > 0.05); There was no significant change in COR at the operated level (P > 0.05). At the last follow-up and at the operated level, COR (Y) showed negative correlations with ROM and translation (P < 0.05), but no follow-up data correlated with COR (X) were found (P > 0.05).ConclusionsSatisfactory clinical and radiological outcomes were achieved 7 years or more after single-level Bryan ACDR. At the operated level, preoperative COR was maintained, probably due to replicating the physiological interrelations of COR (Y), translation, and ROM.