An automated standing type Quick Scan Whole Body Monitor (QS-WBM) has been developed for the measurement of internal radioactive contamination due to high energy photon (HEP) emitters (Eϒ >200keV). Individuals are monitored while standing on a platform inside QS-WBM at specified reference position. Instances may occur where individuals deviate from their monitoring position, potentially leading to errors in the measurement of the body content. Positional inaccuracies are estimated in three potential scenarios: lateral displacement of the individual, movement perpendicular to the detector face, and forward bending of the torso. The Monte Carlo particle transport code FLUKA is used for estimation of efficiencies for selected radionuclides across various geometrical positions for adult male radiation workers. Approximately, 8% variation in calibration factors (CFs) is noted for lateral movements up to 15 cm from the centerline regardless of the energy, while perpendicular movements from reference position exhibit variations ranging from 14 to 60 %. This study helps in quantifying uncertainties in the estimation of body content and dose, arising from deviations in position of individuals in QS-WBM enclosure during monitoring.
Read full abstract