Recently, the ICRU released Report 95, where new operational quantities for external radiation exposure are defined. The new quantities are defined in close relation to the protection quantities. This change affects the practice use of dosemeters. That is why the instruments must be adapted to the measurement of new quantities before their implementing as legally binding. The discrepancies depend on radiation spectra-particle type, energy of particles and direction of incidence. To analyse the performance of currently used instruments, irradiations in photon and neutron fields of various energies were performed for personal and area dosemeters. In this work, the response of photon and neutron personal dosemeters in conditions of rotational geometry is presented. The difference between the responses with respect to the new or old operational quantity was not large, which corresponded to the similar reference values for this irradiation geometry. The mutual ratio depended on the specific radiation quality and geometry. The behaviour of different types of dosemeters varied too.