Medical interventions have significantly progressed in developing minimally invasive techniques like percutaneous procedures. These procedures include biopsy and internal radiation therapy, where a needle or needle-like medical device is inserted through the skin to access a target inside the body. Ensuring accurate needle insertion and minimizing tissue-damage or cracks are critical in these procedures. This research aims to examine the coated needle effect on the force required to insert the needle (i.e., insertion force) and on tissue-damage during needle insertion into the bovine kidney. Reducing the needle insertion force, which is influenced by needle surface friction, generally results in a reduction in tissue-damage. Surgical needles were coated with a composite material, combining Polytetrafluoroethylene, Polydopamine, and Activated Carbon. Force measurement during needle insertion and a histological study to determine tissue-damage were conducted to evaluate the effectiveness of the coating. The insertion force was reduced by 49 % in the case of the coated needles. Furthermore, a histological analysis comparing tissue-damage resulting from coated and uncoated needles revealed an average 39 % reduction in tissue-damage with the use of coated needles. The results of this study demonstrate the potential of coated needles to enhance needle insertion and safety during percutaneous procedures.