Abstract
The development of acute severe mitral regurgitation (MR) requiring surgery is one of the most dreaded complications of percutaneous transmitral commissurotomy. We present the case of a 29 year-old female with severe rheumatic mitral stenosis whose mitral valve dilatation was performed using an Accura balloon. Once the balloon catheter had crossed the mitral valve to reach the left ventricle, it was facing away from the apico-mitral axis. It was inflated in distal portion, and mitral valve dilatation was done in the usual fashion. She developed acute severe MR. As it was misaligned with the apico-mitral axis, it caused a chordae rupture of the postero-medial papillary muscle which was identified during surgery. We conclude that a misaligned Accura balloon catheter during balloon inflation can pluck the chordae tendinae and can cause acute severe mitral regurgitation, and therefore can unearth the underlying subvalvular pathology.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.