Abstract

Here we present an analytical method for Zinc (Zn) isotopic measurements using “standard‐sample bracketing” on a Nu Sapphire multi‐collector inductively coupled plasma‐mass spectrometer. The effects of Zn mass fraction and HNO3 molarity mismatch between the standard and the sample, as well as the presence of matrix elements have been evaluated. Long‐term reproducibility of better than ±0.03‰ for δ66Zn (2s) was routinely obtained. Accurate measurements were achieved when CNi/CZn < 0.001, CTi/CZn < 0.03, CBa/CZn < 0.05, CNa/CZn and CAl/CZn < 0.5, CMg/CZn < 0.1, CFe/CZn < 5 and CCu/CZn as well as CCd/CZn < 10. High‐precision Zn isotopic determination were performed on twenty‐one widely available geological certified reference materials, with an overall range of ~ 0.62‰ (0.15 to 1.07‰), which is nearly twenty times the current analytical precision (0.03), and the results are in agreement with most previously published data within 2s. Among them, δ66ZnJMC 3‐0749L of eleven geological certified reference materials are reported for the first time: 0.22 ± 0.03‰ (dolerite, DR‐N), 0.23 ± 0.02‰ (gabbro, GSR‐10), 0.20 ± 0.04‰ (microgabbro, PM‐S), 0.15 ± 0.02‰ (andesite, GSR‐2), 0.22 ± 0.08‰ (diorite, GSR‐9), 0.19 ± 0.07‰ (syenite, GSR‐7), 0.30 ± 0.06‰ (granite, GSR‐1), 0.25 ± 0.00‰ (rhyolite, GSR‐11), 0.37 ± 0.01‰ (shale, GSR‐5), 0.21 ± 0.03‰ (limestone, GSR‐6) and 0.20 ± 0.03‰ (hornblendite, GSR‐15). The novel Zn isotopic data from these certified reference materials can be used for future interlaboratory comparisons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call