Abstract

Zika virus detection was demonstrated using antibody-functionalized cover glasses externally connected to the gate electrode of an AlGaN/GaN high electron mobility transistor (HEMT). A pulsed bias voltage of 0.5 V was applied to an electrode on the region of the cover glass region functionalized with antibody, and the resulting changes of drain current of the HEMT were employed to determine the presence of Zika virus antigen concentration ranging from 0.1 to 100 ng/ml. The dynamic and static drain current changes as a function of Zika virus concentration were modeled with a spring-like elastic relaxation model and the Langmuir extension model, respectively. Excellent fits to the data were found with relaxation time constants of antibody and antigen molecules in the range of 11 μs and 0.66–24.4 μs, respectively, for the concentration range investigated. The ratio of antibody bound with antigen to the total available antibody on the functionalized contact window was in the range of 0.013–0.84 for the Zika antigen concentration range of 0.1–100 ng/ml. Since the HEMT is not exposed to the bio-solution, it can be used repeatedly. The functionalized glass is the only disposable part in the detection system, showing the potential of this approach for hand-held, low cost sensor packages for point-of-care applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.