Abstract

Radiation damage to macromolecular crystals at third-generation synchrotron sites constitutes a major source of systematic error in X-ray data collection. Here, a computational method to partially correct the observed intensities during data reduction is described and investigated. The method consists of a redundancy-based zero-dose extrapolation of a decay function that is fitted to the intensities of all observations of a unique reflection as a function of dose. It is shown in a test case with weak anomalous signal that this conceptually simple correction, when applied to each unique reflection, can significantly improve the accuracy of averaged intensities and single-wavelength anomalous dispersion phases and leads to enhanced experimental electron-density maps. Limitations of and possible improvements to the method are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.