Abstract

A detailed investigation of the performance as a dosimeter of state-of-art polycrystalline CVD (pCVD) diamond detectors operated in photovoltaic regime for applications in clinical radiotherapy has been carried out with conventional 6–10–18 MV X-photons, as well as with a 10 MV photon Intensity Modulated Radiation Therapy (IMRT) beam from a linear accelerator. Our results show that the performances of a pCVD diamond dosimeter improves dramatically when operated in null-bias conditions. Main improvements with respect to operation with an external voltage applied are: a reduced pre-irradiation dose; an excellent time stability, characterised by standard deviations less than 0.5%; a rise-time comparable to that of commercial reference dosimeters; a linearity with dose proven over three decades; a reduced deviation from linearity of the current vs. dose–rate curve, output factors comparable to that of commercial reference dosimeters. These results represent a significant step towards clinical applications as IMRT with synthetic polycrystalline CVD diamond films.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.