Abstract

Most modern radiology facilities are all digital, and film is not available for physics QC testing. An important test is the measurement of the effective focal spot sizes of the x-ray tube which has an impact on overall image system spatial resolution. Focal spot size measurements typically utilize a star pattern which is magnified and recorded on plain film. Plain film is ideal because it has a spatial resolution of 50-75 LP/mm. Utilizing digital image receptors to record the star pattern images does degrade the measurement because of the size of the individual detector elements (DEL). This article presents a method for the determination of focal spot size using the digital image receptor (DIR) to record the image of the star pattern. An equation which corrects for the image blur introduced by the finite size of the digital detector is given. Measurements of effective focal spot sizes are still important to the assessment of radiographic spatial resolution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.