Abstract
The results of the development of a radiographic installation that allows to control the size of the internal elements of HPGe detectors' cryostats for simulation by Monte Carlo method, are presented. Calibration and assessment of image quality value of the developed radiographic installation were carried out in accordance with ISO Standards 19232–1 and 19232–5. Simulation of the HPGe detector's registration efficiency in the range of 50 keV-3 MeV with the MCC-MT program showed that the discrepancy between the calculated and experimental results when adjusting the detector model using radiological measurement data does not exceed 3% in geometry for point sources and 5% for the Eu-152 source in the Marinelli beaker geometry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.