Abstract

In this work the results on X-ray micro-imaging by means of novel polycapillary optical elements will be presented. To simulate various radiation propagation processes in both single capillary and polycapillary systems, a PolyCAD code was developed. Many experiments have recently revealed the advantage of confocal optical configuration for the fluorescence studies. Moreover, our experimental prototype layout (developped in the INFN - LNF laboratories, Frascati) enables the possibility to obtain μXRF mapping in simultaneous with X-Imaging. The recorded image of an extended sample is limited to 6 μm by the CCD pixel dimensions; the use of a second polycapillary optics in the confocal scheme followed by a SDD detector provides an additional option for elemental studies. A prototype of compact XRF spectrometer with a spatial resolution less than 100 μm has been designed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.