Abstract

An automated wavelength-dispersive x-ray fluorescence spectrometer is operational for analysis of major elements in rocks and minerals. Procedures for trace-element analysis are being developed. Sample preparation methods and analytical techniques are similar to those commonly used elsewhere, but data reduction is conducted by the Fundamental Parameters program developed by Criss. Unlike empirically derived calibration curves, this data reduction method considers x-ray absorption and secondary fluorescence, which vary with differences in sample composition. X-ray intensities for each element from several standards are averaged to develop a theoretical standard for comparison with samples of unknown composition. Accurate data for samples with wide compositional ranges result from these data reduction and standardization techniques.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.