Abstract

A photon-counting K-edge x-ray Computed Tomography (CT) system is useful for discriminating photon energy and for decreasing absorbed dose for patients. The CT system is of the first generation type and consists of an x-ray generator, a turn table, a translation stage, a two-stage controller, a multipixel photon counter (MPPC) module, a 0.5-mm-thick zinc oxide (ZnO) scintillator, a counter board (CB), and a personal computer (PC). Tomography is accomplished by repeating the translation and rotation of an object. Penetrating x-ray photons from the object are detected by the scintillator in conjunction with the MPPC module, and the event signals are counted by the CB. Without using energy discriminating, photon counting CT was carried out by controlling x-ray spectra.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.