Abstract

Radiochromic film (RC-film) is of great interest as a film-type dosimeter for radiation oncology applications. We present a two-dimensional image-based evaluation of the measurement accuracy of a commercial RC-film product (Gafchromic MD-55-2 film, ISP TECHNOLOGIES, Inc.) by using a commercial Laser Densitometer (Model 1710, Computerized Medical Systems, Inc.) as an optical density imaging system. The coefficient of variation of the density (pixel-value) in one sample was approximately 3% to 11% at 3 Gy or less, and 3% or less at 4 to 60 Gy. Although the coefficient of variation between three samples at the same dose was about 14% at 1 Gy, it decreased as the dose increased, reaching several percent. In 1 to 6 Gy samples, geometric imaging artifacts [interference (moire) patterns] were observed, and it was found that scan-sampling pitch influenced the accuracy of measurement of the density of the sample. To improve the accuracy of density measurement, sufficient knowledge about characteristic features of the density measuring system is essential.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.