Abstract

The relative calibration of a chemical vapor deposited (CVD) diamond X-ray detector has been completed on the shenguang-Ⅱ laser facility. Signals of the CVD diamond detector and an absolute-calibrated flat response X-ray diode (XRD) for X-ray radiation were obtained. The average sensitivity of the CVD diamond X-ray detector is about 1.19610-5 C/J. The difference between average sensitivity and the sensitivities in different shots is no more than 13%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.