The ultra-thin 2in. diameter silicon (111)-H(1×1) wafer appeared to be a promising material for a mirror focusing He-atom beam in a scanning atom microscope. To increase achievable at present resolution from 1.5μm to a sub-micron range the 50μm thick (111) silicon wafer with thickness distribution better than ±1μm and 0.05° precision of miscut value is necessary. The purpose of this paper was to adapt X-ray measurements to control the miscut value with high precision for the 50μm thick silicon wafers as well as to control the deviation of the wafer surface from ideal flat plane by measurements of low-angle reflection using X-ray high resolution diffractometer (HRXRD). The crucial point was the construction of a stress-free ultra-thin wafer holder. The precision of 0.01° miscut value was obtained. The deviation of the surface from the ideal flat surface obtained by X-ray measurements were compared with the one estimated from optical (confocal) method. A satisfactory good conformity between both methods has been observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.