Abstract

Implementation of the recommendations of the new version of the ISO 4037 standard at X-ray installations of dosimetry calibration laboratories proved to be difficult and the uncertainties on the conversion coefficients due to differences between the fluence spectra from different installations are not yet sufficiently established. The X-ray installation of SCK CEN/LNK was characterized to verify the conformity with the most recent version of the ISO 4037 standard. The Half-Value Layer (HVL) values were measured to prove that conditions for matched beam qualities are fulfilled and then photon spectrometry with a High Purity Germanium (HPGe) detector was performed to measure the fluence spectra. Using the fluence spectra, the conversion coefficients from air kerma Ka to operational quantities were calculated and compared with the recommended values from the ISO 4037 standard. Most of these coefficients agree within 2% while the ISO 4037 recommended values have an uncertainty of 4% (k = 2). Possible sources of uncertainty in the photon spectrometry were evaluated and are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.