Abstract

The pulse height spectra of the radiation qualities N-10 to N-120 of the ISO narrow spectrum series (N) were measured using a commercial Ge spectrometer and a specially designed lead collimator. For the unfolding of these spectra, three response matrices of the Ge spectrometer for different maximum energies and energy resolutions were calculated by the EGS4 (Electron Gamma Shower version 4) code incorporating the PRESTA and LSCAT options. Then, the pulse height spectra were unfolded with two different algorithms to check the correctness. To verify the measuring and unfolding method, the first half value layers (HVLs) were determined using small ionisation chambers and the fluence spectra. The mean photon energies were calculated, also on the basis of the fluence spectra, and compared with published values. For each radiation quality N-10 to N-120 the conversion coefficients hpK(10;N,a) from air kerma Ka to personal dose equivalent at 10mm depth, Hp (10,a), were calculated for angles of incidence a of 15, 30, 45, 60(and 75(between the unidirectional photon field and the normal to the slab phantom surface. Because the fluence spectra were influenced by the air density, in particular the low energy spectra, the hpK(10;N,a) values were normalised to reference conditions. Differences of up to about 88% between the hpK(10;N,a) values presented in this work and those given in ISO/FDIS 4037-3 were found.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.