Abstract
Semiconductor hybrid pixel detectors, originally developed for particle physics experiments, have been used for an X-ray diffraction experiment on a synchrotron radiation source. The spatial resolution of the intensity peaks in the diffraction patterns of silicon and potassium niobate powder samples was found to be better than that of a scintillator-based system, typically used at present. The two-dimensional position information of the pixel detector enabled multi-peak diffraction patterns to be acquired and clearly resolved without the need for an angle scan with a diffractometer. This trial experiment shows the potential of this technology for high-resolution high-rate diffraction systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.