Abstract

We have developed a MWPC system for X-ray detection with a position resolution in the region of 6 μm rms (14 μm fwhm). The performance with argon, krypton or xenon at pressures from 1 to 10 atm is explored for the X-ray energy range 5–25 keV. At a resolution of 6 μm rms the effects of photoelectron and Auger electron range, electronic noise, avalanche spread, lateral electron diffusion, as well as X-ray beam collimation, become of comparable magnitude. Their limiting effects on avalanche centroid fluctuation, and hence on position resolution, are investigated. The position resolution achieved in this work compares favorably with that of solid state devices.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.