Abstract

We report here the results of a mesh experiment to measure the subpixel structure of the EPIC MOS CCDs on board the XMM X-ray observatory. The pixel size is 40 μ m square while the mesh hole spacing is 48 μ m, a combination quite different from our standard mesh experiment. We have verified that this combination functions properly and have analyzed the CCD structure with sub-pixel resolution. The EPIC MOS CCD has an open electrode structure to improve detection efficiency at low energies. We obtained the distribution of various grades of X-ray events inside the pixel. A horizontally split two-pixel event is generated near the channel stop which forms a straight vertical pixel boundary whereas a vertically split two-pixel event is generated where the potential due to the thinned gate structure forms a wavy horizontal pixel boundary. Therefore, the effective pixel shape is not a square but is distorted. The distribution of X-ray events clearly shows that the two etched regions in each pixel, separated by the bridging finger of the enlarged (open) electrode. We measured the difference in X-ray transmission between the conventional and open regions of the pixel using O–K and Cu–L X-ray emission lines, and found it to be consistent with an electrode thickness comprising 0.2±0.1 μ m of Si and 0.6±0.2 μ m of SiO 2.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.