Abstract
The objective of this study is to develop and test a unique X-ray-induced acoustic computed tomography system that combines the advantages of high X-ray imaging contrast and high ultrasonic spatial resolution. The system features a 5 MHz 128-element ultrasound transducer ring-array formed into a full circular aperture. A parallel data receiver, which consists of a dedicated 128-channel preamplifier and a 128-channel data acquisition module, provides full tomographic imaging at a speed of up to 25 frames per second. Details of the system design and calibration are presented, along with the characteristic results of the imaging resolution. The tomographic imaging performance is demonstrated through images of a phantom with a spatial resolution up to 138 μm. The study results indicate that this imaging device and the methodology provide a rapid and high resolution approach for the dynamic imaging of information, and it may have the potential for becoming a promising noninvasive imaging modality to be used in future applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.